| Earth/Ground Tester |
|
|
| Specifications |
|
| ||
Operation: | Central rotary switch and function keys | ||
Working temperature range: | … +50 °C | ||
Operating temperature range: | 0 | °C | … +35 °C |
Nominal temperature range: | 18 | °C | … +28 °C |
Storage temperature range: | … +60 °C |
Note
The chart of four temperature ranges for the instrument exist to sat- isfy European Standards requirements; the instrument can be used over the full Working temperature range by using the temperature co- efficient to calculate accuracy at the ambient temperature of use.
Temperature coefficient: | ± 0.1 % of range / Kelvin |
Operating errors: | refer to operating temperature range and RH |
| < 20 RE, RS < 100 RE |
The maximum percentage operating error within the measurement range does not exceed ± 30 % with the measured value as fiducial value, as determined in accordance with TABLE 1.
The operating error applies under the rated operating conditions given in
•injection of series interference voltages with system frequencies of 400 Hz, 60 Hz, 50 Hz, 16 2/3 Hz or with d.c. voltage respectively across the terminals E (ES) and S. The r.m.s. value of the series inter- ference voltage shall be 3 V;
•resistance of the auxiliary earth electrode and of the probes: 0 to 100 x RA but ≤ 50 kΩ;
•system voltages between 85 % and 110 % of the nominal voltage and between 99 % and 101 % of the nominal system frequency for meas- uring equipment with a mains supply and/or measuring equipment de- riving its output voltage directly from the distribution system.
Limits of error: | refer to nominal temperature range |
Climate class: | C1 (IEC |
Type of protection: | IP 56 for case, IP 40 for battery door according to |
| EN 60529 |
11