Fluke 1625 user manual Earth/Ground Tester, Specifications

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 19
Image 19

 

 

 

Earth/Ground Tester

 

 

 

Specifications

 

 

Operation:

Central rotary switch and function keys

Working temperature range:

-10 °C

… +50 °C

Operating temperature range:

0

°C

… +35 °C

Nominal temperature range:

18

°C

… +28 °C

Storage temperature range:

-30 °C

… +60 °C

Note

The chart of four temperature ranges for the instrument exist to sat- isfy European Standards requirements; the instrument can be used over the full Working temperature range by using the temperature co- efficient to calculate accuracy at the ambient temperature of use.

Temperature coefficient:

± 0.1 % of range / Kelvin

Operating errors:

refer to operating temperature range and RH

 

< 20 RE, RS < 100 RE

The maximum percentage operating error within the measurement range does not exceed ± 30 % with the measured value as fiducial value, as determined in accordance with TABLE 1.

The operating error applies under the rated operating conditions given in IEC1557-1 and the following:

injection of series interference voltages with system frequencies of 400 Hz, 60 Hz, 50 Hz, 16 2/3 Hz or with d.c. voltage respectively across the terminals E (ES) and S. The r.m.s. value of the series inter- ference voltage shall be 3 V;

resistance of the auxiliary earth electrode and of the probes: 0 to 100 x RA but ≤ 50 kΩ;

system voltages between 85 % and 110 % of the nominal voltage and between 99 % and 101 % of the nominal system frequency for meas- uring equipment with a mains supply and/or measuring equipment de- riving its output voltage directly from the distribution system.

Limits of error:

refer to nominal temperature range

Climate class:

C1 (IEC 654-1), -5 °C...+45 °C, 5 %...95 % RH

Type of protection:

IP 56 for case, IP 40 for battery door according to

 

EN 60529

11

Page 19
Image 19
Fluke 1625 user manual Earth/Ground Tester, Specifications