Fluke 1625 R E withclip − ontransformer, R E withoutclip − onTransfprmer, trnew = trold ⋅

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 51
Image 51
trnew = trold ⋅

Earth/Ground Tester

Procedure of Measurements

 

1625 EARTH / GROUND TESTER

START

 

 

H/C2

TEST

 

 

 

DISPLAY

 

 

S/P2

MENU

 

 

 

 

AC

 

 

Earth/Ground Resistance 300 k

Resistance

DC Low Resistance 3k

 

300 k

 

RA

R

R

 

4 POLE

2 POLE

 

 

 

2 POLE

 

3 POLE

 

 

4 POLE

 

4 POLE

CHANGE

3 POLE

 

ES/P1

ITEM

 

 

 

OFF

 

 

SELECT

 

 

E/C1

 

 

 

edw017.eps

Figure 14. Correcting Clip-on Transformer Errors

2.Turn central rotary switch to position "ARE 3pole".

3.Press "START TEST"-button and note result of RE value.

4.Connect clip-on transformer.

 

1625 EARTH / GROUND TESTER

START

 

 

H/C2

TEST

 

 

 

DISPLAY

 

 

S/P2

MENU

 

 

 

 

 

 

 

AC

 

 

Earth/Ground Resistance 300 k

Resistance

DC Low Resistance 3k

 

300 k

 

RA

R

R

 

4 POLE

2 POLE

 

 

 

2 POLE

 

3 POLE

 

 

4 POLE

 

4 POLE

CHANGE

3 POLE

 

ES/P1

ITEM

 

 

 

OFF

 

 

SELECT

 

 

E/C1

 

 

 

edw018.eps

5.Turn central rotary switch to position "ARE 3pole".

6.Press "START TEST" again.

If the thus measured value RE deviates from the RE value determined without clip-on transformer by more than 5 %, adjust the clip-on transformation ratio (tr) correspondingly:

trnew = trold

RE (withclip ontransformer)

RE (withoutclip onTransfprmer)

43

Page 51
Image 51
Fluke 1625 user manual R E withclip − ontransformer, R E withoutclip − onTransfprmer, trnew = trold ⋅, Earth/Ground Tester