Earth/Ground Tester
Procedure of Measurements
Safety Control Measurements
Before each measurement the instrument automatically checks the measuring conditions and, while simultaneously displaying the kind of error, prevents measurements from being started under the following conditions:
•excessive voltage on the sockets (> 24 V in RE and R~; > 3 V in RF)
•wrong or incomplete connection
•Problems during the measuring sequence (display "E1 ... E5") see display description in the section ‘Procedure of Measurements’.
•Battery voltage too low ( display
Measurement of Interference - Voltages and Frequen- cies
This measuring function detects possible interference voltages and their fre- quencies. This function is automatically active in every switch position before an earthing or resistance measurement. If the
Bring central rotary switch in desired position, read out measured value of in- terference voltage, measured value of interference frequency is displayed with "DISPLAY".
31