Fluke 1625 user manual Resistance Measurement RF, Earth/Ground Tester, Specifications

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 27
Image 27

 

 

 

 

 

 

 

Earth/Ground Tester

 

 

 

 

 

 

 

 

Specifications

 

 

 

 

 

 

 

 

Resistance Measurement (RF)

 

 

 

Measuring method:

current- voltage measurement as per IEC61557-4

 

 

 

 

 

possible

 

 

 

 

Open circuit voltage:

20 V DC

 

 

 

 

Short circuit current:

200 mA DC

 

 

 

Formation of measured

with 4-pole measurement wires on H, S, ES can

 

value:

 

be extended without additional error.

 

 

 

 

 

 

Resistances >1 Ω in wire E can cause additional

 

 

 

 

 

error of 5m Ω/Ω.

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring

 

Display Range

 

Resolution

Intrinsic

Operating

 

 

Range

 

 

 

 

 

Error

Error

 

 

 

 

 

 

 

 

 

 

 

 

 

0.001 ... 2.999

 

0.001

 

 

 

 

 

 

 

 

 

 

 

 

 

0.020

 

3.0 ... 29.99

 

0.01

±( 2 % of

±( 5% of

 

 

 

 

 

 

 

 

m.v. +2

m.v. +5

 

 

3 k

 

30.0 ... 299.9

 

0.1

 

 

 

 

digit)

digit)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

300 ...

2999

 

1

 

 

 

 

 

 

 

 

 

 

 

Measuring sequence:

approx. 2 measurements/s

 

 

Measuring time:

 

typ. 4 sec. incl. reversal of polarity (2 pole or 4

 

 

 

 

 

pole)

 

 

 

 

Max. interference voltage:

≤ 3 V AC or DC, with higher voltages measure-

 

 

 

 

 

ment will not be started

 

 

 

Max inductivity:

 

2 Henry

 

 

 

 

Max. overload:

 

Urms = 250 V

 

 

 