Fluke 1625 user manual Principle of Operation, Purpose, Stakeless Earth/Ground Resistance Testing

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 73
Image 73

 

Stakeless Earth/Ground Resistance Testing

 

Principle of Operation

 

 

Dimensions

Total length of adaptercable 1,8 m, Length

 

of shielded cable to clamp-on CT 1,5 m.

Operating error*:

± (10% of m.v. + 5 digit).

Display range*:

0,010 ... 130 Ω.

Measuring frequency:

128 Hz.

Measuring voltage:

Um=48 VAC (primary).

Measuring range*:

0,020 ... 100 Ω.

Resolution:

.001... .1 ohm.

*when using:

CT EI-162AC to induce voltage and

 

CT EI-162X to detect current.

 

Recommended distance between CT's:

 

10 cm.

Principle of Operation

Purpose

Testing the resistances of individual ground connections in systems with paral- lel ground connections (multi-grounded systems).

65

Page 73
Image 73