Resistance Measurement (RF)

Earth/Ground Tester

Procedure of Measurements

2.Connect instrument according to picture.

3.In this mode, all settings and LIMIT values available can be called with "DISPLAY MENU" and the measuring frequency can be set.

4.Press "START TEST" button.

5.Read out measured value.

Resistance Measurement (RF)

In this measuring mode all resistances from 0.001 Ω to 3 kΩ can be measured with DC voltage and automatic polarity reversal as per EN61557-5.

To achieve highest accuracy 4 pole measurements are possible. To balance the extension lead, a compensation has to be done.

 

1625 EARTH / GROUND TESTER

START

 

 

H/C2

TEST

 

 

 

DISPLAY

 

 

S/P2

MENU

 

 

 

 

AC

 

 

Earth/Ground Resistance 300 k

Resistance

DC Low Resistance 3k

 

300 k

 

RA

R

R

 

4 POLE

2 POLE

 

 

 

2 POLE

 

3 POLE

 

 

4 POLE

 

4 POLE

CHANGE

3 POLE

 

ES/P1

ITEM

 

 

 

OFF

 

 

SELECT

 

 

E/C1

 

 

 

edw024.eps

Figure 18. Resistance Measurement (RF)

1.Connect instrument according to picture.

2.Turn central rotary switch to position "RF".

3.In this mode, all settings and LIMIT values available can be called with "DISPLAY MENU".

WWarning

Before starting a measurement bring plant or test object to off or de-energized circuit condition! With an external volt- age higher than 3 V measurement will not be started.