Fluke 1625 user manual Measurement of Interference Voltage DC + AC UST, Earthing Resistance R E

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 21
Image 21
Measurement of Interference Voltage DC + AC (UST)

Earth/Ground Tester

Specifications

Measurement of Interference Voltage DC + AC (UST)

Measuring method:

fullwave rectification

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring

 

Display

 

Resolu-

Frequency

 

Limits of Er-

 

Range

 

Range

 

tion

 

Range

 

 

ror

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1…50 V

 

0.0…50 V

 

0.1 V

 

DC/AC

 

 

± (5% of

 

 

 

 

 

 

 

 

45…400 Hz

 

reading +5

 

 

 

 

 

 

 

 

sine

 

 

digit)

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring sequence:

approx. 4 measurements /s

 

 

 

Internal resistance:

approx. 1.5 MΩ

 

 

 

 

 

Max. overload:

 

Urms = 250 V

 

 

 

 

 

Measurement of Interference Frequency (FST)

 

Measuring method:

Measurement of oscillation period of the interfer-

 

 

 

 

ence voltage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measuring

Display Range

 

Resolution

Range

 

Limits of

 

Range

 

 

Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16.0 … 400

16.0…299.9…99

 

0.1 … 1 Hz

1 V …

 

± (1% v. mv

 

Hz

9 Hz

 

 

 

 

 

50 V

 

+2 digit)

 

 

 

 

 

 

 

 

 

 

 

Earthing Resistance (RE)

 

 

 

 

 

 

 

Measuring method:

 

current and voltage measurement with probe

 

 

 

 

 

as IEC61557-5

 

 

 

 

 

Open circuit voltage:

 

20 / 48 V, AC

 

 

 

 

 

Short circuit current:

 

250 mA

AC

 

 

 

 

 

Measuring frequency:

 

94 , 105 , 111 , 128 Hz selected manually or

 

 

 

 

 

automatic.(AFC) 55 Hz in function R*

 

Noise rejection:

 

 

120 dB ( 16 2/3, 50 , 60, 400 Hz )

 

Max. overload:

 

 

Urms = 250 V

 

 

 

 

 

13

Page 21
Image 21
Fluke 1625 Measurement of Interference Voltage DC + AC UST, Measurement of Interference Frequency F ST, Specifications