| Earth/Ground Tester | ||
|
|
|
|
|
|
|
|
| Specifications | ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| A | = | intrinsic error | B[%] = ± |
| B |
| x 100% | ||
|
| En |
| = | variations | fiducial value | |||||
|
| R | = | routine test |
|
|
|
|
| ||
|
| T | = | type test |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| |
|
| Measuring |
|
|
|
| Instrinsic | Max. Op- | |||
|
| Display Range |
| Resolution |
|
| erating | ||||
|
| Range |
|
| Error |
| |||||
|
|
|
|
|
|
| Error | ||||
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 0.001 Ω...2.999 Ω |
| 0.001 Ω |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 3.00 Ω...29.99 Ω |
| 0.01 Ω |
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
| 0.020 Ω | 30.0 Ω...299.9 Ω |
| 0.1 Ω |
| ± ( 2 % of |
| ± ( 5% of | ||
|
|
|
|
|
|
|
|
| mv +2 digit ) |
| mv +5 |
|
| 300 kΩ |
| 0.300 kΩ...2.999 kΩ |
| 1 Ω |
|
| |||
|
|
|
|
|
|
| digit ) | ||||
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 3.00 kΩ...29.99 kΩ |
| 10 Ω |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 30.0 kΩ...299.9 kΩ |
| 100 Ω |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Measuring time:
Additional error because of
Measuring error of RH and RS:
Max. probe resistance:
Max. auxiliary earth elec- trode resistance:
typ. 8 sec. with a fixed frequency
30 sec. max. with AFC and complete cycle of all
measuring frequencies
RH (RS + 2000Ω) ⋅1.25 ⋅10−6% + 5digits
RE
typ. 10 % of RE + Rs + RH
≤1 M Ω
≤1 M Ω
Automatic check if error is kept within the limits required by
If after a measurement of
15