Fluke 1625 user manual Users Manual, cordance with, Parts of, Test

Models: 1625

1 90
Download 90 pages 25.79 Kb
Page 22
Image 22
cordance with

1625

Users Manual

Table 2. Electrical Measurement Specifications

 

 

 

Requirements

 

Intrinsic Er-

Reference Con-

 

or Test in Ac-

Type

ror or Influ-

ditions or Speci-

Designation

cordance with

of

ence Quan-

fied Operating

Code

the Relevant

Test

tity

Range

 

Parts of

 

 

 

 

 

IEC 1557

 

 

 

 

 

 

Intrinsic error

Reference condi-

A

Part 5, 6.1

R

 

tions

 

 

 

 

 

 

 

 

Position

Reference posi-

E1

Part 1, 4.2

R

 

tion

 

 

 

 

± 90°

 

 

 

 

 

 

 

 

Supply volt-

At the limits

E2

Part 1, 4.2,

R

age

stated by the

 

4.3

 

 

manufacturer

 

 

 

 

 

 

 

 

Temperature

0 °C and 35 °C

E3

Part 1, 4.2

T

 

 

 

 

 

Series inter-

See 4.2 and 4.3

E4

Part 5, 4.2,

T

ference volt-

 

 

4.3

 

age

 

 

 

 

 

 

 

 

 

Resistance

0 to 100 x RA

E5

Part 5, 4.3

T

of the probes

but 50 k

 

 

 

and auxiliary

 

 

 

 

 

 

 

earth elec-

 

 

 

 

trodes

 

 

 

 

 

 

 

 

 

System fre-

99 % to 101 % of

E7

Part 5, 4.3

T

quency

the nominal fre-

 

 

 

 

quency

 

 

 

 

 

 

 

 

System volt-

85 % to 110 % of

E8

Part 5, 4.3

T

age

the nominal volt-

 

 

 

 

age

 

 

 

 

 

 

 

 

Operating

 

 

Part 5, 4.3

R

error

B = ±( A + 1,15 E12 E22 E32 E42 E52 E62 E72 E82

 

 

 

 

 

 

 

14

Page 22
Image 22
Fluke 1625 user manual Users Manual, cordance with, Parts of, Test