Power Recorder

Navigating in Power View on the PDA

Icons on the Menu Bar

and are used to toggle between voltage and current wave forms. Tap either V or A and then Live>Home to see each parameter’s detailed data. The PHASOR screen always shows both V & A (volts and amps). is replaced with on the detail views, as voltage and current data can be viewed simultaneously.

In all the detail live view screens, a line selection menu is available using the button at the bottom. This is a checkbox that is used to select the voltage and current channels you want to display.

The Snapshot icon triggers the Recorder to capture high-resolution waveform data and save it in the recorded data stream. This waveform capture is identical to the snapshots taken automatically when you set the “snapshot period” in the 1750 Setup menu.

This icon brings up the Annotate screen. Use the annotate menu to enter text or to insert a file name and load it into the recorded data stream in the Recorder. Annotations can be used to make notes, or to insert a file name of an image or audio file. You can enter multiple start and stop annotations. START and stop annotations are not paired and are not restricted to being used as literally “start” and “stop”. Annotations are often done at the start of a power recording session.