3654 Portable H2/N2 Analyzer - Specifications

55 of 66

 

 

7

Specifications

 

 

7.1 Instrument Specifications

 

 

 

Table 7-1:General Technical Data

 

 

 

 

 

 

 

Power Requirements

Batteries: two C-type cells, NiCad or alkaline, each 26 mm

 

 

x 50 mm, 2.4 to 3 volts total

 

 

 

 

 

Battery Autonomy

15 hours operational use, 3 weeks purge backup (power

 

 

off, purge gas supplied)

 

 

 

 

 

Dimensions (HxWxD)

19.8 x 11.5 x 22 cm

 

 

Weight (including purge gas

4.6 kg

 

 

cylinder and sensor)

 

 

 

 

 

Purge Gas

Nitrogen (H2 sensors)

 

 

Carbon Dioxide (N2 and H2 sensors)

 

 

 

 

 

Purge Gas Cylinder Pressure

200 bar maximum for N2 purge gas

 

 

Limit (see also Note below)

60 bar maximum for CO2 purge gas

 

 

 

40 hours when cylinder filled to 200 bar of N2

 

 

Purge Gas Autonomy

20 hours when cylinder filled to 100 bar of N2

 

 

 

120 hours when cylinder filled with 75% of CO2 liquid

 

 

Capacity of Purge Gas Cylinder

65 ml

 

 

 

Baud rate: 4800

 

 

 

Data bits: 8

 

 

Digital Output (RS232C)

Stop Bits: 1

 

 

 

Start Bits: 0

 

 

 

Parity: None

 

 

Measurement Cycle Time

20 seconds

 

 

Temperature of Use

0 to 40°C (32 to 104°F)

 

 

Temperature of Sample

-5 to 35°C (23 to 95°F)

 

 

Sample Pressure Limit (32013A

< 10 bar, at <= 25° C

 

 

and 32015 flow chambers)

 

 

 

Enclosure Protection

IP 67

 

 

CE Certification

EN 61326-1:1997 /A1:1998 /A2:2001 /A3:2003

 

 

Directive 89/336/CE

 

 

 

 

 

 

 

Note:

In accordance with the Pressure Equipment Directive 97/23/EC (Article 3.3), the components that make up the purge gas cylinder and pressure reducer, are not subject to the technical requirements of this directive.

Operator Manual

ORBISPHERE

Page 61
Image 61
Hach 3654 Portable H2/N2 Analyzer manual Instrument Specifications