3654 Portable H2/N2 Analyzer - Options Setup 33 of 66
Operator Manual ORBISPHERE
3.4.8 Automatic Shutdown3.4.9 Measurement Mode
Choose Measurement Mode from the Configuration menu to enable maximum
measurement mode for sampling in bottles and cans.
When this occurs, the display is frozen and you may store the data. The instrument
remains in this mode until you select Normal measurement mode.
Choose Automatic Shutdown Status from the Configuration
menu to activate the Automatic Shutdown feature.
If you select Enable, the instrument switches off automatically
after 10 minutes of inactivity, thereby economizing battery
power.
In Maximum measurement mode, the
instrument searches for two consecutive
measurements with less than a 2%
difference, as also described in
“Measurement Mode” on page 19.