54 of 66Maintenance and Troubleshooting - 3654 Portable H2/N2 Analyzer

6.6.4 Troubleshooting Table

Table 6-2:Troubleshooting Table

Symptom

Cause

Possible Solution

 

Degassing

Adjust sample flow rate

 

 

 

 

Flow chamber and/or membrane not clean

Clean system

 

 

 

 

Sensor’s membrane not tight smooth and

Replace membrane and

 

wrinkle free

recalibrate

 

 

 

Unstable

 

No gas bubbles should escape

measurement

 

from the purge gas outlet during

 

Leaking solenoid valve

measurement. If the solenoid

 

leaks (more than 1 bubble in 15

 

 

 

 

seconds) contact your Hach Ultra

 

 

representative.

 

 

 

 

An external power supply is used but the

Switch to battery power

 

power supply is not clean

 

 

 

 

 

 

Sample flow rate too low

Check flow rate

 

 

 

 

Degassing

Adjust forcing gas and/or sample

 

flow rate

 

 

 

 

 

 

 

Allow sample to flow past the

 

 

membrane to make sure the

Low readings

Membrane is not at sample temperature

membrane is at sample

 

 

temperature before making

 

 

measurements

 

 

 

 

Membrane not clean

Clean system

 

 

 

 

Incorrect solubility curve

Change solubility

 

 

 

 

Incorrect calibration

Recalibrate

 

 

 

 

 

Allow sample to flow past the

 

 

membrane to make sure the

 

Membrane is not at sample temperature

membrane is at sample

High readings

 

temperature before making

 

measurements

 

 

 

 

 

 

Incorrect solubility curve

Change solubility

 

 

 

 

Incorrect calibration

Recalibrate

 

 

 

ORBISPHERE

Operator Manual

Page 60
Image 60
Hach 3654 Portable H2/N2 Analyzer manual Troubleshooting Table, Symptom Cause Possible Solution