56 of 66Specifications - 3654 Portable H2/N2 Analyzer

7.2 Sensor Specifications

Table 7-2: Sensor Specifications - Table 1

Sensor Type

 

31 270 H2

 

 

 

 

 

Membranes

29561A

2952A

2935A

 

 

 

 

Thickness

25 µm

25 µm

25 µm

 

 

 

 

Membrane material

PFA

ETFE

ECTFE

(Halar)

 

 

 

 

 

 

 

Recommended

Waste gas offgas,

Reactor

High H2 level

applications

reactor coolant

coolant

 

 

 

 

Radiation limits

105 rad

108 rad

108 rad

Measurement range at

0-1.5 ppm, or

0-5 ppm, or

0-15 ppm, or

0-16 cc/kg, or

0-60 cc/kg, or

0-180 cc/kg, or

25°C

0-1 bar

0-3 bar

0-9 bar

 

 

 

 

 

Accuracy (sample temp.

The greater of

The greater of

The greater of

±1% of reading

±1% of reading

±1% of reading

20-50°C within

or ± 4 ppb,

or ± 16 ppb,

or ± 45 ppb,

± 5°C of calibration

or ± 0.06cc/kg,

or ± 0.2cc/kg,

or ± 0.5 cc/kg,

temperature)

or ± 3 mbar

or ± 12 mbar

or ± 30 mbar

 

 

 

 

 

 

The greater of

The greater of

The greater of

Accuracy (sample temp.

±2% of reading or

±2% of reading

±2% of reading

0-50°C independent of

± 30 ppb or

or ± 120 ppb

or ± 300 ppb

calibration temperature)

± 0.35 cc/kg or

or ± 1.2 cc/kg,

or ± 3.0 cc/kg,

 

± 12 mbar

or ± 40 mbar

or ± 100 mbar

 

 

 

 

Measurement cycle

 

20

 

(sec.)

 

 

 

 

 

 

 

 

 

Sample flow rate

200 ml/min

100 ml/min

50 ml/min

(through flow chamber)

 

 

 

 

 

 

 

ORBISPHERE

Operator Manual

Page 62
Image 62
Hach 3654 Portable H2/N2 Analyzer manual Sensor Specifications, Sensor Type 31 270 H