Chapter 2, Making Measurements

Check Even-Second Clock

Check Even-Second Clock

Overview

Most CDMA transmitter tests require an Even-Second clock signal from the base

 

station in order to correlate the measurement. If you are receiving an error stating

 

Cannot correlate to input signal. Check setup. , you may have a

 

problem with your Even-Second clock signal. This Software utility allows you to

 

check for a valid clock signal from the base station.

EVEN SECOND/

(Rear Panel)

SYNC IN

 

H

 

 

SYNTH

Even-Second

REF IN

 

Clock

 

Time Base

 

H

 

Figure 24

 

 

 

NOTE:

The test procedures outlined in this section require that you have already configured the

 

Test System as outlined in chapter 1, "Getting Started".

 

 

Requirements

• This test requires that you temporarily connect the Even-Second clock signal to the

 

TRIGGER / QUALIFIER IN port of the Cellular Adapter.

 

 

NOTE:

Be sure to return the Even-Second clock signal to the EVEN SECOND / SYNC IN port of

 

the Cellular Adapter when finished. The Software will prompt you to make this change.

 

 