Model 5328A

Maintenance

 

 

Table 5-8A. Program Description

Check

Test

Observe on 5328A

Point

 

 

 

1

REMOTE

Front panel (RMT) annunciator should be on.

2

CHECK

Counter should read 10.000 MHz.

3

RESOLUTION

The 9825A should print and counter display the 10 MHz

 

 

check signal with resolutions from 0.1 Hz to 1.0 MHz.

4

FREQ A

Counter display should read 0.0000 KHz.

5

RATlO B/A

Counter display should read 0.0000000

6

PERIOD A

Counter display should read 0. s

7

PER. AVE. A

Counter display should read 0.00000 ns

8

T.I. A-B

Counter display should read 0. s

9

T.I. AVG. A-B

Counter display should read 0.00000 ns

10

FREQ C

Counter display should read 0.0000 KHz

11

RATIO C/A

Counter display should read 0.0000000

12

SAMPLE RATE

When calculator displays MANUAL OK?, verify that front

 

SINGLE/

panel SAMPLE RATE control can be manually adjusted as

 

MULTIPLE

seen from GATE LIGHT flashing rate.

 

MEASMNT

When calculator displays GATE LIGHT OFF?, verify that Gate

 

 

Light is truly off.

 

 

 

5-42. Connect a function generator to the 5328 input channels and monitor the signal with an oscilloscope as shown in Figure 5-8.

Figure 5-8.Hookup for tests described in Table 5-8B & C

5 - 3 6