3-47. SLOPE SWITCH. The ±SLOPE switch (provided for each channel) determines which slope of the input signal will trigger the counter. As a simple example, (Figure 3-9) if the pulse width of a positive pulse is to be measured, the A channel slope switch would be set to “+” and the B chan- nel would be set to “-” (for time interval measurements the A channel always begins the mea- surement and the B channel ends the measurement).

N O T E

A simple pulse width measurement is achieved with the use of the +SLOPE setting for Channel A and the -SLOPE setting for Channel B.

Figure 3–9.Slope Switch Settings

3-48. LEVEL CONTROL. The LEVEL control for each channel is adjustable over the range of ±2.5V dc with the attenuator for that channel in the X1 position. A typical use of the LEVEL con- trols is shown in Figure 3-10.

N O T E

Simple measurement of a time interval, the LEVEL control of the A and B input channels were used to set the trigger LEVEL of A and B.

Figure 3-10.Level Control Settings