Model 5328A

Maintenance

Table 5-4. Performance Test (Continued)

. Set the pulse generator for the following output:

a.(-) SLOPE TEST

Execute the following from the 9825A keyboard:

wrt 701, “PF8G1S13A57+000*B57+000*R”

Counter should display 20 µS ±10 µs. Mark results on performance test record.

b.(+) SLOPE TEST

. Press RESET button on HP 5328A. Set Channels A and B slope switches to (-).

. Execute the following from the 9825A keyboard.

wrt 701, “PF8G1S13A7+000*B7+000*R”

5-12

Page 79
Image 79
HP E42 manual Performance Test