Model 5328A
Maintenance
c.Set 5328A rear panel control as follows:
ARM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
d.Connect HP 608E Signal Generator (or equivalent) to INPUT A. Set signal gen- erator to 35 MHz at 50 mV rms (140 mV
e.Slowly decrease the signal generators output level to 15 mV rms (42 mV
f.To set Channel B sensitivity change 5328A front panel controls as follows:
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RATIO B/A
RESOLUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1(33 10 kHz
LEVEL B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PRESET
g.With HP 651B Test Oscillator (set to 10 MHz at 100 mV rms) connected to lN- PUT A, connect a second signal generator (set to 40 MHz at 50 mV rms) to iN- PUT B.
h.Repeat step e adjusting variable resistor R28 instead of R26.
2.Channel C Sensitivity adjustments:
a.Remove the top cover from the 5328A.
b.Set signal to 100 MHz and reduce level until no stable reading in counter dis- play. Adjust A8R82 for stable reading.
c.Repeat step b. until best sensitivity is obtained.
d.Repeat step b. to ensure that the counter still meets the requirement.
3.High Frequency Offset adjustments:
a.Remove top cover.
b.Set signal generator to 500 MHz and reduce signal level until display reading is no longer stable. Adjust A8R85 until display is stable.
c.Repeat step b. until best balance is obtained.
4.
The following adjustment procedure adjusts the All
The procedure offsets an input signal to the 5328A by 0, +2, and