Model 5328A

Maintenance

Table 5-4. Performance Test (Continued]

Connect the 5328A A MARKER output to the A channel of the oscilloscope. Adjust the position of the A MARKER as described above and verify that it intersects the 100 Hz sinewave at +2 volts.

Execute the following from the 9825A keyboard:

wrt 701, “PF4G6S13A379-200*B37-200*R“

Adjust the display of the A channel marker output such that the top of the wave- form just barely intersects both positive and negative slopes of the 100 Hz waveform. Verify that this occurs at -2 volts on the 100 Hz waveform.

Connect the 5328A B marker output to the A channel of the oscilloscope. Adjust the position of the B marker as described and verify that it intersects the 100 Hz wave- form at -2 volts.

Mark results on performance test record.

5-17

Page 84
Image 84
HP E42 manual Performance Test