Model 5328A

Maintenance

 

 

 

Table 5-8B. Program Description

Check

Step

Test

Counter Display Readout

Point

131 ATTENUATOR Set the function generator to an output of 1 kHz, 100 mV p-p sinewave centered at @vdc as seen on the oscilloscope. Connect the function

 

 

 

 

generator’s output to Channel A of the counter. Set the counter’s

 

 

 

 

LEVEL A and B to PRESET. HP5328AF/096/H42 trigger lights should

 

 

 

 

be blinking.

 

2

 

 

When the 9825A CONTINUE key is pressed, verify that the

 

 

 

 

counter trigger lights stop blinking.

14

3

SEPARATE/

 

With function generator connected to Channel A of counter,

 

 

COMMON

 

when CONTINUE key of 9825A is pressed, verify counter readout

 

 

NORMAL/

 

as 1.000.

 

4

INVERTED

 

With function generator connected to Channel B of counter, when

 

 

 

 

 

 

 

CONTINUE key of 9325A is pressed, counter should display

 

 

 

 

approximately 1.00 kHz (frequency of function generator).

15

5

COUPLING

 

Set the function generator to a triangular pulse output of 1 kHz at

 

 

 

 

300 mV p-p with a +0.4 vdc offset (Figure 5-9).Connect signal to

 

 

 

 

counter’s Channel A input. Counter’s Channel A and B trigger lights

 

 

 

 

should be blinking.

 

6

 

 

When the 9825A CONTINUE key is pressed, observe the counter’s

 

 

 

 

trigger lights stop blinking.

16

7

SLOPE/

 

Set the function generator to a triangular pulse output of 1 kHz at

 

 

TRIGGER

 

300 mV p-p with a +0.4 vdc offset (Figure 5-9).Connect the function

 

 

LEVEL

 

generator’s output to Channel A of the counter. Set the counter’s

 

 

 

 

LEVEL A and B to PRESET.

 

8

 

 

When the 9825A CONTINUE key is pressed, observe Channel A

 

 

 

 

and B trigger lights commence blinking.

 

9

Channel

B

When the 9825A CONTINUE key is pressed, counter should

 

 

+ SLOPE

 

display approximately 150µsec±75µsec (wide tolerance).

 

10

Channel

B

When the 9825A CONTINUE key is pressed, counter should

 

 

- SLOPE

 

display approximately 400µsec±100µsec (wide tolerance).

 

11

Channel

A

When the 9825A CONTINUE key is pressed, counter should

 

 

+ SLOPE

 

display approximately 400µsec±100µsec (wide tolerance).

 

12

Channel

A

When the 9825A CONTINUE key is pressed, counter should

 

 

- SLOPE

 

display approximately 150µsec±75µsec (wide tolerance).

5-37

Page 108
Image 108