Leupold 21 PRO manual Trouble shooting Búsqueda de fallos

Models: 21 PRO 21

1 24
Download 24 pages 32.52 Kb
Page 14
Image 14
Trouble shooting • Búsqueda de fallos

 

Trouble shooting • Búsqueda de fallos

english

 

Recherche de panne • Localização de avarias

 

 

Power

No LED lighting up

Possible causes:

 

 

 

TEST

Ningún LED está iluminado

• Power supply faulty

 

• Power supply board defective

 

 

 

Valve

Alarm

Aucune LED n'est allumée

• Control PCB defective

 

 

- Check voltage on type plate.

 

 

Todos os LEDs apagados

 

 

- Check wiring (external and internal)

Alarm

 

 

- Check plug connections

Valve

TEST

 

- Check printed circuit boards for

 

possible damage

Power

 

 

 

 

 

 

 

 

 

Power

Pressing of test button, but no

Possible causes:

condensate discharge

• Feed and/or outlet line shut off or

 

 

 

TEST

 

 

El interruptor de „TEST“ está

blocked

 

 

• Worn parts (seals, valve core,

Valve

Alarm

pulsado, pero el condensado no

diaphragm)

 

 

se evacua

 

 

• Control PCB defective

 

 

 

 

 

La touche Test est actionnée,

• Solenoid valve defective

Alarm

 

• Dropping below necessary minimum

 

mais sans purge du condensat

 

pressure

Valve

TEST

 

Botão de teste premido, mas não

- Check feed line and outlet line

Power

 

 

há descarga de condensado

- Replace worn parts

 

 

 

 

 

- Check if valve opens audibly (press

 

 

 

test button several times)

 

 

 

- Check printed circuit boards for

 

 

 

possible damage

 

 

 

- Check operating pressure; where

 

 

 

necessary, install pressure or

 

 

 

vacuum drains.

Power

 

Condensate discharge only when

Possible causes:

 

test button is being pressed

• Feed line with insufficient slope;

 

 

 

 

 

 

TEST

Evacuación del condensado soló

crosssection too small.

 

 

 

• Excessive condensate quantities

Valve

Alarm

 

si está pulsado el interruptor de

 

• Sensor tube extremely dirty

 

 

 

„TEST“

 

 

 

- Lay feed line with adequate slope

 

 

 

 

Alarm

 

 

Purge du condensat uniquement si

- Install venting line

 

 

la touche Test est actionnée

- Clean sensor tube

Valve

 

TEST

 

 

 

Power

 

 

Descarga de condensado só com

 

 

 

 

o botão de teste premido

 

Power

 

 

Device keeps blowing off air

Possible causes:

 

 

 

• Control air line blocked

 

 

 

El aparato está abierto

 

 

TEST

• Worn parts (seals, valve core,

 

 

constantemente

 

Alarm

 

diaphragm)

Valve

 

 

 

L'appareil refoule de l'air en

 

 

 

 

- Clean entire valve unit

 

 

 

permanence

- Replace worn parts

Alarm

 

 

O purgador perde ar

- Clean sensor tube

 

 

 

 

 

 

 

Valve

 

TEST

continuamente

 

Power

 

 

 

 

14

BEKOMAT 21, 21 PRO

Page 14
Image 14
Leupold 21 PRO manual Trouble shooting Búsqueda de fallos