How to do a control solution test:

STEP 1 Insert Test Strip.

Insert a test strip, contact bars end first and facing up, into the test port. Push it all the way in until it will go no further.

The meter will turn on and the display check will appear briefly.

Then the code number will appear, followed by the R symbol and unit of measure. Check that the unit of measure is set correctly. Be sure the meter and test strip codes match. If they do not, code the meter correctly.

Press the C button to mark the test as a control solution test in the meter memory. Ë will appear on the display. If you decide not to do a control solution test, press the C button again to remove Ë from the display.