Table 23. Backup and Target Rates
FeatureDescription
Backup Rate

Rate that the CRT-P uses to pace when

 

there is a sudden rate decrease. This can

 

be a maximum of 10 ppm less than the

 

intrinsic rate and follows the Target Rate with

 

a 1 to 6 ppm per cycle increase or

 

0.25…1.25 ppm per cycle If the Target Rate

 

is less than the current Back Up Rate.

Target Rate

The Target Rate is either the current

 

detection rate minus 10 ppm, or the sensor

 

or basic rate. The Backup Rate follows the

 

Target Rate with the programmed rate

 

increase or decrease.

Figure 8: Rate Fading

2.17 Home Monitoring (Stratos LV-T)

Home Monitoring enables the exchange of information about a patient’s cardiac status from the implant to the physician. Home Monitoring can be used to provide the physician with advance reports from the implant and can process them into graphical and tabular format called a Cardio Report. This information helps the physician optimize the therapy process, as it allows the patient to be scheduled for additional clinical appointments between regular follow-up visits if necessary.