100 Stratos
Feature | Description |
Backup Rate | Rate that the |
| there is a sudden rate decrease. This can |
| be a maximum of 10 ppm less than the |
| intrinsic rate and follows the Target Rate with |
| a 1 to 6 ppm per cycle increase or |
| 0.25…1.25 ppm per cycle If the Target Rate |
| is less than the current Back Up Rate. |
Target Rate | The Target Rate is either the current |
| detection rate minus 10 ppm, or the sensor |
| or basic rate. The Backup Rate follows the |
| Target Rate with the programmed rate |
| increase or decrease. |
2.17 Home Monitoring (Stratos LV-T)
Home Monitoring enables the exchange of information about a patient’s cardiac status from the implant to the physician. Home Monitoring can be used to provide the physician with advance reports from the implant and can process them into graphical and tabular format called a Cardio Report. This information helps the physician optimize the therapy process, as it allows the patient to be scheduled for additional clinical appointments between regular