Stratos
When the suspicion criterion has been met, the Stratos
The upper interval limit range must be shorter than the limit of the VA delay (350 ms, for example). The test method is based on the length of the pacing interval or the AV delay (refer to Table 22).
Table 22. PMT Test MethodInterval Length | AV Delay | Test Method |
> UTI (upper | ≤ 200 ms | Increasing the AV |
tracking interval) |
| delay by 50 ms |
> UTI + 50 ms | > 200 ms | Reducing the AV |
|
| delay by 50 ms |
≤ UTI | ≤ 200 ms | Increasing the UTI |
|
| by 50 ms |
≤ UTI | > 200 ms | Increasing the UTI |
|
| by 50 ms |
> UTI and ≤ UTI + | > 200 ms | Length of UTI = TA |
50 ms |
| + 50 ms |
Stratos
2.10 Adjustment of the PMT Protection Window
The PMT protection window can be automatically adjusted. This automatic adjustment functions in the following manner: