5.4.5 NIPS Safety Features
The BIOTRONIK offers the following safety features during NIPS sessions.
•Before the NIPS feature can be used, NIPS must be specifically selected and then is released through user acknowledgment. In addition, before NIPS is performed in the Ventricle, the user must acknowledge that delivering NIPS into the Ventricle may induce dangerous arrhythmias.
•When the battery voltage has reached the Elective Replacement Indicator point (ERI), the NIPS feature is no longer available.
•Ventricular pacing support is available to
•NIPS may only be programmed temporarily.
NOTE:
High pacing rates and pulse amplitudes together with wide pulse widths may temporarily decrease the amplitude of the pacing pulse. The pacing pulse must be continuously verified with an ECG to assure effectiveness.
To perform NIPS function, the programmer wand must be placed directly over the