148 Stratos LV/LV-T Technical Manual

Table 32 shows the mean* expected time intervals (in months) from ERI to EOS at standard program for Stratos CRT-Ps. All service intervals, including the above-cited nominal longevity, are based on the battery discharge behavior and the hybrid circuit properties including current consumption and replacement indicator. The statistical calculations are based on 500 ohm lead impedances, 100% pacing, and data supplied by the battery manufacturer.

Table 32: Remaining Expected Service time (ERI to EOS)
Pacing ProgramStratos LV, (months)

Standard, Mean

8

 

 

Program with high pulse energy

8

 

 

*50% of all pacemakers reach or exceed the given value

Standard: DDDR, 60 ppm, 3.6 V, 0.4 ms

High: DDDR, 90 ppm, 4.8 V, 1.0 ms