Stratos LV/LV-T Technical Manual 143

1.Insert the enclosed torque wrench through the perforation in the self- sealing plug at an angle perpendicular to the lead connector until it is firmly placed in the setscrew.

2.Insert the lead connector pin into the connector receptacle of the CRT-P without bending the lead until the connector pin becomes visible behind the setscrew. Hold the connector in this position.

3.Securely tighten the setscrew of the connector clockwise with the torque wrench until torque transmission is limited by the wrench.

4.After retracting the torque wrench, the perforation will self-seal. The proximal electrode of bipolar leads is automatically connected. Connect the second lead as described above.

5.Attach the other ventricular lead to the right ventricular port (RV) and the atrial lead to the right atrial port (RA) in the same manner.