Stratos LV/LV-T Technical Manual 145
9. Elective Replacement Indication (ERI)Service times for the Stratos CRT-Ps vary based on several factors, including battery properties, storage time, lead system impedance, programmed parameters, amount of pacing and sensing required, and circuit operating characteristics. Service time is the time from beginning of service (BOS) to the end of service (EOS). To assist the physician in determining the optimum time for pulse generator replacement, an elective replacement indicator is provided that is activated when the battery cell capacity drops to a predetermined level. The following table defines the different service cycles (at standard settings, 37°C, and with a lead impedance of 500 ohms). The beginning of the replacement cycle is displayed on the programmer after pulse generator interrogation and appears on the printout.
Table 29: Service Cycle DefinitionsAbbreviation | Service | Definition |
| Cycle | |
BOS | Beginning of | Normal service cycle; |
| Service | battery in good condition |
ERI | Elective | Identifies the time of |
| Replacement | elective replacement |
| Indication | indication. The rate |
| | occurring at ERI depends |
| | upon the programmed |
| | mode and magnet |
| | application. |
EOS | End of | Identifies the end of the |
| Service | elective replacement |
| | indication period. |
The Stratos CRT-Ps indicate the need for replacement by a defined decrease in the programmed pacing rate without a magnet applied. The rate change is dependent on the programmed pacing mode.