Stratos
Rate Fading Rate Increase
0.5; 1; 2… (1)…6 ppm/cycle
Rate Fading Rate Decrease
0.25; 0.5… (0.25)…1.25 ppm/cycle
Lead Configuration
Pacing for the Stratos LV /
UNIP; BIPL (3 separate channels)
Sensing for the Stratos LV /
UNIP; BIPL (2 separate channels; 3 IEGM channels)
Lead check
Off; On (3 separate channels)
Rate Adaptation
Off; On
Sensor
Accelerometer
Sensor Gain
Auto; 1…40 (in 32 steps)
Automatic Sensor Gain
Off; On
Sensor Threshold
Very low; low; medium, high; very high
Rate Increase
0.5; 1… (1)…6 ppm/cycle
Rate Decrease
0.25… (0.25)…1.25 ppm/cycle
Maximum Sensor Rate
90… (5)…120… (5)…180 ppm