Stratos LV/LV-T Technical Manual 107

NOTE:

The attending physician must notify the BIOTRONIK Service Center about which of these events he/she wishes to be informed.

2.17.4.3 Patient Report

It is possible to trigger a transmission through magnet application over the CRT-P. The attending physician must inform the patient in detail about operating the device and about the physical symptoms which would warrant a magnet application by the patient.

CAUTION

Patient-Activated Report - The magnet effect must be programmed “synchronous” if the [Patient Report] function is activated.

2.17.5 Description of Transmitted Data

The following data are transmitted by the Home Monitoring system, when activated. In addition to the medical data, the serial number of the implant is also transmitted.

The Monitoring Interval

Type of Last Home Monitoring Message and Time of the Transmission of the last Home Monitoring Message provide information regarding the interval of Home Monitoring data.

Resynchronization Therapy

% Pacing in the Atrium / 24 hours [%]

Atrial intrinsic rhythm [%]

CRT - %ventricular pacing [%]

Ventricular paces (AV Delay expired) [%]

Ventricular paces (triggered by RV Sense) [%]