The length of the time interval (monitoring interval) is preset to “daily”. For each monitoring interval, a data set is generated in the implant and the transmission is initiated at the designated time.
2.17.4.2 Event ReportWhen certain cardiac and technical events are detected by the implant, a report transmission is automatically triggered. This is described as an “event message”.
The following cardiac and technical events initiate a message transmission:
• | Atrial | Lead | Check | • | Low | Amplitude*) | |
| < 300 and > 3000 Ohm |
| (<50% safety margin) | ||||
• | Ventricular Lead | Check | • | Low | Amplitude* | ||
| < 300 and > 3000 Ohm |
| (<50% safety margin) | ||||
• | ACC Disabled |
| • | Ventricular Threshold >4.8V |
CAUTION
Event Triggered Report - A timely receipt of the event report cannot be guaranteed. The receipt is also dependent on whether the patient was physically situated in the required coverage range of the patient device at the time the event information was sent.
*) Examples: The programmed sensitivity is 1.0 mV.