152 Stratos LV/LV-T Technical Manual

Dynamic AV Delay (Dual chamber only)

low; medium; high; individual; fixed

AV Delay Values (Dual chamber modes only)

15… (5)…300

AV Hysteresis

Off; 10… (10)…100 ms

Repetitive AV Hysteresis

Off; 1… (1)…10

Scan AV Hysteresis

Off; 1… (1)…10

Sense Compensation

Off; -10… (-10)…-120 ms

Safety AV Delay

100 ms

Ventricular Blanking Time

30… (5)…70 ms

Magnet effect

Automatic; Auto; asynchronous; synchronous Asynchronous Magnet Effect: paces at 90 ppm.

Automatic Magnet Effect: 10 cycles at 90 ppm asynchronous; thereafter synchronous with the programmed basic rate

Synchronous Magnet Effect: synchronous with programmed basic rate

High Rate Protection

205…215 ppm

Pulse Amplitude

0.2... (0.1)...3.6... (0.1)...6.2; 7.2 V

(3 channels separately programmable)