152 Stratos
Dynamic AV Delay (Dual chamber only)
low; medium; high; individual; fixed
AV Delay Values (Dual chamber modes only)
15… (5)…300
AV Hysteresis
Off; 10… (10)…100 ms
Repetitive AV Hysteresis
Off; 1… (1)…10
Scan AV Hysteresis
Off; 1… (1)…10
Sense Compensation
Off;
Safety AV Delay
100 ms
Ventricular Blanking Time
30… (5)…70 ms
Magnet effect
Automatic; Auto; asynchronous; synchronous Asynchronous Magnet Effect: paces at 90 ppm.
Automatic Magnet Effect: 10 cycles at 90 ppm asynchronous; thereafter synchronous with the programmed basic rate
Synchronous Magnet Effect: synchronous with programmed basic rate
High Rate Protection
205…215 ppm
Pulse Amplitude
0.2... (0.1)...3.6... (0.1)...6.2; 7.2 V
(3 channels separately programmable)