8 Stratos
Lead Configuration – The polarity of the implanted lead dictates what lead configuration can be programmed for the
Setscrew Adjustment –
Cross Threading Setscrew(s) – To prevent cross threading the setscrew(s), do not back the setscrew(s) completely out of the threaded hole. Leave the torque wrench in the slot of the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s). Use only the BIOTRONIK supplied torque wrench.
Sealing System – Be sure to properly insert the torque wrench into the perforation at an angle perpendicular to the connector receptacle. Failure to do so may result in damage to the plug and its
1.5.4 Programming and Operation
IEGM – Due to the compression processes that the signals undergo, the IEGM recordings are not suitable for making some specific cardiac diagnoses, such as ischemia; although, these tracings may be useful in diagnosing arrhythmias, device behavior or programming issues.
Post AES - Before activating
Overdrive Pacing Mode - When programming the overdrive pacing mode, check whether the selected program can cause PMT, and whether atrial over drive pacing would result. Corresponding to the measured retrograde conduction time, the PMT protection interval must be programmed to a correct value.