130 Stratos LV/LV-T Technical Manual

6.2Patient Data Memory

Individual patient data can be stored in the Stratos CRT-Ps. For example, the following are stored:

Patient name

Patient index (how the patient name is coded)

Implantation date

Symptoms

Etiology

ECG indication

QRS width

LV ejection fraction

Lead polarity

6.3Safe Program Settings

Activating the preset values for the Safe Program is a quick and convenient way to provide VVI / SSI pacing at a high output setting in urgent situations. Listed in Table 26 are the Safe Program settings for Stratos CRT-Ps.

Table 26: Safe Program Settings

Parameter

Safe Program Settings

Mode

VVI

Pacing Rate

70 ppm

Amplitude

4.8 V (ventricle)

Pulse Width

1.0 ms

Sensitivity

2.5 mV

Ventricular Refractory Period

300 ms

Pacing Polarity

Unipolar

Sensing Polarity

Unipolar

Single Chamber Hysteresis

OFF

Magnet Effect

AUTO