Stratos LV/LV-T Technical Manual 159

11.7 Default Programs

Stratos LV / LV-T

Parameter/Factory settings/

Safe Program

FunctionStandard Program

 

Mode

DDD

VVI

 

 

 

VV

BiV RV RV-T

-

Synchronization

 

 

AV delay

150 ms at 60 ppm

120ms at 130

 

 

ppm

First Chamber

RV

-

Paced

 

 

VV delay after

5 ms

-

pace

 

 

VV delay after

5 ms

-

sense

 

 

Pulse

3.6 V (A); 3.6 V

4.8 V

Amplitude

(RV); 3.6 V (LV)

 

Sensing A/RV

UNIP

UNIP (RV)

 

 

 

Refractory

Auto (A); 250 ms (V)

A: N/A

Period

 

V: 300 ms

11.8 Materials in Contact with Human Tissue

Housing: Titanium

Connector receptacle: Epoxy resin

Sealing Plugs: Silicone Rubber

11.9 Electrical Data/Battery

NOTE:

At 37º C, with pacing impedance of 500 Ohms.