154 Stratos LV/LV-T Technical Manual

Intervention Rate

100… (10)…250 ppm

X-out-of-8 Activation Criterion

3… (1)…8

Z-out-of-8 Termination Criterion

3… (1)…8

DDI(R) Basic Rate

32… (1)…88… (2)…122… (3)…140… (5)…180 ppm

Overdrive Modes:

Off; On

Available in the modes DDD, DDTA, AAI, AAT with and without rate adaptation

Maximum Overdrive Pacing Rate

90… (5)…160 ppm

Overdrive Pacing Increment (Rate Increase)

2… (2)…10 ppm

Overpacing Level (Rate Drop After)

1… (1) …32 cycles

Post-AES Pacing

Off; On

AES Increment

5… (5)…40 ppm

AES Prematurity

5…50%

VES discrimination after As

OFF; 250… (50)…450 ms

Rate Fading

Off; On