Stratos
2.1.11 OFF (ODO)
In this mode, pacing and sensing functions are off. The OFF mode is used to determine and evaluate the morphology of an intrinsic rhythm. With external pulse control, the OFF mode is used for electrophysiological studies. The OFF mode can be programmed temporarily.
CAUTION
OFF Mode – The OFF mode can be transmitted as a temporary program only to permit evaluation of the patient’s spontaneous rhythm. (see Section 2.1.11).
2.2Biventricular Synchronization of the
Stratos CRT-Ps
For the Stratos
OFF - If the BiV sync parameter is set to OFF, the
BiV RV
The biventricular synch settings can be activated together with the DDD(R), DDI(R), VDD(R), and VVI(R) pacing modes.
CAUTION
Sensing – The Stratos
AV Conduction – In patients with intact AV conduction, the intrinsic atrial tachycardia is conducted to the ventricle 1:1. With the resynchronization mode activated, spontaneous rate of the right ventricle mode is synchronized for a rate up to 200 ppm in the left ventricle. For this reason, biventricular pacing mode should be turned OFF in such cases.