The following tables summarizes what parameters you can change.

TABLE 3-1 Range Parameters

 

Option

F/C

Resolution

Range

S1

S4

S5

 

 

*-MF1

¡F

-328 to + 1562¡F

--

A

A

 

 

*-MF2

¡F

0.1¡

-199.9 to + 199.9¡F

C

--

C

 

 

*-MC1

¡C

-200 to + 850¡C

--

A

B

 

 

*-MC2

¡C

0.1¡

-199.9 to + 199.9¡C

C

--

B

 

 

 

 

 

 

 

 

 

 

* Refer to Table 1-1, for complete models (preface section).

TABLE 3-2 2-, 3-, 4-Wire, Upscale, Downscale Configuration

 

Configuration

S2

S3

 

 

S3

 

 

 

 

 

 

Upscale (Default)

 

Downscale

 

 

2- or 4-wire RTD

--

--

 

A, B, E, F

 

C, D, G, H

 

 

3-Wire RTD

A

I

 

A, B, E, F

 

C, D, G, H

 

 

 

 

 

 

 

 

 

 

Note 3-Wire input connection is standard. For 2- or 4-Wire remove S2-A and S3-I