3.6 Display Hold Connections

3.6 Display Hold Connections

The ÒDisplay Hold" function can be used to freeze the display. Connect a single pole, single throw (SPST) switch between pins 1 and 2 on the P2 connector. The display will stop updating when this switch is on. When this switch is off, the display will update itself 2.5 times per second.

To access the ÒDisplay Hold" function, attach copper wire to the 3-pin female connector that is supplied with the meter. Refer to Figure 3-10for wiring instructions. This connector shares a common return with the Analog Output. Plug the female connector into the P2 connector at the rear of the meter.

Figure 2-2shows the location of the P2 connector on the meter.

Switch

Display Hold

RETURN

1 Manual background 2

Screws

3

P2

underneath

Figure 3-10. Display Hold Connections