Omega DP119-RTD Calibration Values When You Are Using Resistance Source, Meter, Pot To, Option

Models: DP119-RTD

1 36
Download 36 pages 21.74 Kb
Page 29
Image 29
CALIBRATION VALUES WHEN YOU ARE USING

P3

4

RED

 

 

+ Excitation

 

RED

 

 

 

Signal Hi

3

 

 

 

BLACK

Signal Lo

2

 

- Excitation

1

BLACK

 

 

 

RED PRECISION RESISTANCE

BLACK SOURCE

Figure 4-3. 4-Wire Configuration

TABLE 4-1

CALIBRATION VALUES WHEN YOU ARE USING

RESISTANCE SOURCE

 

MODEL

RESISTANCE

METER

POT TO

 

 

OPTION

READING

DISPLAY

ADJUST

 

 

*-MF1

93.01

00¡F

ZERO

 

 

 

375.51

1472¡F

SPAN

 

 

*-MF2

93.01

00.0¡F

ZERO

 

 

 

134.70

194.0¡F

SPAN

 

 

*-MC1

100

00¡C

ZERO

 

 

 

375.51

800¡C

SPAN

 

 

*-MC2

100.0¡C

00¡C

ZERO

 

 

 

174.0¡C

195.0¡C

SPAN

 

* Refer to Table 1-1, for complete models (preface section).

ZERO SPAN

Figure 4-4. Calibration Potentiometers

22

Page 29
Image 29
Omega DP119-RTD Calibration Values When You Are Using Resistance Source, Meter, Pot To, Option, Reading, Display, Adjust