SECTION 5 SPECIFICATIONS

INPUT TYPE:

 

 

 

 

 

 

RTD:

 

100-ohm platinum with alpha of 0.00385

Calibration:

 

IEC 751 (IPTS-68)

Configuration:

 

2-, 3-, or 4-wire RTD connection

Polarity:

 

Bipolar

Zero:

 

Adjustable ±20¡C (±36¡F)

Zero Tempco :

 

±0/05 degree/degree

Sensor-wire Resistance Tempco Effect per Conductor:

2-wire:

 

20 mdeg/½/degree, up to 3.8 ½

3-wire:

 

50µdeg/½/degree, up to 100 ½

4-wire:

 

5.0µdeg/½/degree, up to 100 ½

Sensor-excitation Current:

 

0.5 mA

Sensor-break Indication:

 

Meter displays overrange

 

 

 

 

 

 

(three least significant digits blank)

Overvoltage Protection:

 

25 Vp

 

 

 

 

 

 

NOISE REJECTION

 

 

 

NMR , SIG HI to SIG LO:

 

70 dB, 50/60 Hz

CMR , Analog RTN to PWR GND:

120 dB, DC to 60 Hz

CMV , Analog RTN to PWR GND:

Dielectric strength to 1500 V

 

 

 

 

transient per 260 Volt rms or DC working voltage.

ACCURACY at 25¡C

 

 

 

 

Range for rated accuracy:

 

 

 

MC1/MF1:

 

-200¡ to +850¡C (-328¡ to +1562¡F)

Maximum Error:

 

±1.5¡C (±2.7¡F) ±1 count

MC2/MF2:

 

-199.9¡ to +199.9¡C (-199.9¡ to +199.9¡F)

Maximum Error:

 

±0.5¡C (±0.5¡F) ±1 count

Resolution:

0.1¡C, 0.1¡F, 1¡C or 1¡F (jumper-selectable)

Span Tempco :

 

±0.02% of reading/¡C

Full-Scale Step Response:

 

1 second

Warmup to Rated Accuracy:

.............................................................

30 minutes

 

 

 

ANALOG OUTPUT (linearized)

 

Voltage:

 

1mV/count

Calibration Error:

 

 

 

MC1/MF1:

±1% of reading ± 1.5 mV on ¡C, ±2.7mV on ¡F

MC2/MF2:

 

±1% of reading ± 0.5 mV

Source Resistance:

 

100 ½

ANALOG TO DIGITAL CONVERSION

 

Technique:

 

Dual-slope, average value

Signal Integration Period:

 

100 ms

Read Rate:

 

2.5/s