3.8 Disassembly/Assembly

Note

1.

2.

You must go through the following steps in order to remove the printed circuit board from the case if you want to change from 115VAC to 230VAC (or vice versa) or to change from Fahrenheit to Celsius (or vice versa). These changes must be performed by a qualified technician in order to avoid damage to the unit.

WARNING: Do not connect ac power to your meter until you have completed all input and output connections. Failure to do so may result in injury!

Remove power and all wire connections from the unit.

Unplug the three connectors (P1, P2, and P3) from the back of the meter. Grasp them firmly on the top and bottom of the connector and pull. Figure 3-13shows the location of the three connectors.

3.Remove the friction fit lens by prying down and forward from the top of the meter. Remove the mounting screws from the rear of the panel backplate. Slide the unit out of the panel (if it is installed in a panel).

4.The printed circuit board is held in place by a tab and notch system located next to the P1 connector, pin 1 (refer to Figure 3-13). The tab is on the case and the notch is on the edge of the PC Board. Hold the meter with the display facing away from you. Gently pry the tab (on the back corner of the case) out with the edge of your finger, while pushing the back of the PC Board out through the front of the meter. The PC Board will slide out the front, free from the case.

18

Page 25
Image 25
Omega DP119-RTD manual Disassembly/Assembly