OMEGA Model CYD201/CYD208 User’s Manual
F0 | Sets Sensor Units for Temperature Display. |
Input: F0x
Returned: Nothing
Remarks: Sets sensor units for the temperature display, where x = C (Celsius), F (Fahrenheit), K (kelvin), or V (volts).
HSelect High Alarm Setpoint Value.
Input: Hxxx.x
Returned: Nothing
Remarks: Sets high alarm setpoint, where xxx.x = temperature setpoint in units specified by F0 command. If the instrument is set for volts, the alarm defaults to kelvin.
Example: H300[term] sets a high alarm setpoint of 300 degrees.
LSelect Low Alarm Setpoint Value.
Input: Lxxx.x
Returned: Nothing
Remarks: Sets low alarm setpoint, where xxx.x = temperature setpoint in units specified by F0 command. If the instrument is set for volts, the alarm defaults to kelvin.
Example: L31.2[term] sets a low alarm setpoint of 31.2 degrees.
RAlarm Reset.
Input: R
Returned: Nothing
Remarks: Resets the alarm.
YChannel Dwell Time (Model CYD208 Only).
Input: Yab
Returned: Nothing
Remarks: Sets dwell time for a given channel, where a = channel 1 - 8, and b = the dwell time parameter as follows:
0 = zero seconds, 1 = 5 seconds, 2 = 10 seconds, 3 = 30 seconds, 4 = 60 seconds
Setting a dwell time of 0 skips the specified channel in the sequence.
Example: Y23[term] sets the dwell time for channel 2 to 30 seconds.