OMEGA Model CYD201/CYD208 User’s Manual

3.6.2Verifying SoftCal™ Operation

Check the status of SoftCal™ by holding UNITS for 2-3 seconds. If SoftCal™ is not in operation, "-000-" displays.

If the first digit of the display is 1, SoftCal™ is set for below 28 K. If either the second or third digit is 1, SoftCal™ is set for above 28 K.

3.6.3Erasing SoftCal™ Compensations

When SoftCal™ compensations are erased, the thermometer returns to normal operation.

1.Move the SET switch to position 1.

2.Hold UNITS until the front panel displays 0. This key is a toggle. If the display moves in the wrong direction, release and press again.

3.Press CAL ENABLE on rear panel using a pen tip. The display changes from 0 to "-SOF-". The alarm setpoint displays after erasing SoftCal™ compensation.

4.Move the SET switch to position 0.