OMEGA Model CYD201/CYD208 User’s Manual
5.7RECALIBRATION
OMEGA calibrates and certifies thermometers to original factory specifications for a reasonable fee. You can recalibrate the thermometer to original specifications, but OMEGA will not warrant these calibrations.
Recalibration requires a digital voltmeter (DVM) with 4½ digit resolution or better; and 25 kand 125 kprecision resistors with ±0.01% tolerance or better.
Recalibration involves current source and analog/digital (A/D) converter calibration. A/D calibration erases all SoftCal™ compensations; perform it after current source calibration, not before.
5.7.1Current Source Calibration
1.Allow 30 minute
2.Configure 125 kresistor as shown in Figure
3.Connect DVM voltage leads across the resistor and adjust the I ADJ trimpot (located on rear panel) until DVM displays a voltage of 1.2500 volts ±100 microvolts.
4.Remove DVM and resistor.
Model CYD201 | Model CYD208 |
NOTE: Do not use a voltmeter for A/D Converter Calibration.
Figure 5-4. Calibration Connections