OMEGA Model CYD201/CYD208 User’s Manual

5.7.2A/D Converter Calibration

NOTE: Current source calibration must be performed before A/D converter calibration.

1.Allow 30 minute warm-up to achieve rated specifications.

2.Perform current source calibration.

3.Configure 125 k￿resistor as shown in Figure 5-4. Set Model 208s to channel 1.

4.Wait 10 seconds for resistor voltage to settle.

5.Press CAL ENABLE, then within 2 seconds press UNITS. The display window shows "-CAL-" for approximately 15 seconds.

6.Repeat procedure with 25 k￿resistor.

7.For Model CYD208, repeat procedure for other 7 channels. See Figure 2-2 for pin assignments.