B) Function mode |
|
|
|
|
|
|
Setting range | : | 00~33 |
|
|
|
|
Display range | : | 00~39 |
|
|
|
|
|
| 00, 04, 08, 13, 15, 16, 18, 19, 20, 26, 27, 31 and 33 are unavailable. | ||||
Setting procedure : |
| In PV display, keep pressing | key over 5 seconds to enter function | |||
|
| mode (F00 is displayed). Change the blinking 1st digit to desired function | ||||
|
| code with | key and |
| key. Press SET key to be function code | |
|
| available. |
|
|
|
|
|
| If you press SET key after input digits among 00, 04, 08, 13, 15, 16, 18, | ||||
|
| 19, 20, 31 and 33, the display automatically reverts to PV. | ||||
Out of setting range : | If you try to input F34~F39, it is unacceptable to change with | |||||
|
| automatically revert to PV display. |
|
5.Warning function
| High temp. alarm | : | When PV is higher than SV+SVH (high temp. alarm SV) +1, ALARM lamp | |||
|
|
| and LED display blinks, intermittent buzzer beeps with approx. 15 | |||
|
|
| minutes of delay, and remote alarm output turns on. | |||
|
|
| When PV is lower than SV+ SVH, ALARM lamp and LED display go off, | |||
|
|
| buzzer stops beeping, and remote alarm output turns off. | |||
|
|
| If you press BUZZER key, the buzzer stops beeping instead remote | |||
|
|
| alarm output does not turn off. |
| ||
| Low temp. alarm | : | When PV is lower than | |||
|
|
| and LED display blinks, intermittent buzzer beeps with approx. 15 | |||
|
|
| minutes of delay, and remote alarm output turns on. | |||
|
|
| When PV is lower than SV- SVL, ALARM lamp and LED display go off, | |||
|
|
| buzzer stops beeping, and remote alarm output turns off. | |||
|
|
| If you press BUZZER key, the buzzer stops beeping instead remote | |||
|
|
| alarm output does not turn off. |
| ||
6. | Other function |
|
|
|
|
|
| Auto Return | : | If there is not any key operation for 90 seconds in each setting mode, | |||
|
|
| Key Lock mode and Function mode, the display automatically reverts to | |||
|
|
| PV. Note) Auto Return is not worked in F09 and F10. | |||
| Sensor temperature : F12: Temperature in temp. sensor is displayed |
| ||||
|
|
| (Ex. | 㩷 | 㩷 displayed as ‘80.2’) |
|
|
|
| F14: Temperature of comp. sensor is displayed. |
| ||
|
|
| (Ex. 67 㩷 | 㩷 | displayed as ‘067’) |
|
| Battery accumulating | F03: Battery accumulating time is displayed. |
| |||
| time | : | (Ex. 2years and 6months consumed 㩷 displayed as ‘02.5’) | |||
|
|
| The display shows ’02.8’ to inform the battery exhaustion. | |||
|
|
| Then replace with new battery. |
| ||
|
|
| <Reset of accumulating time> |
| ||
|
|
| Step into F06. Input ‘409’ and press SET key to turn the display | |||
| ROM version | : | to ’00.0’ with BATTERY lamp goes off. |
| ||
| F30: ROM version is displayed (Ex. Ver. 1.00 | 㩷 displayed as “1.00”) | ||||
| Condensing fan |
| F32: Notice for condensing fan motor accumulating time | |||
| motor accumulating |
| 000: Not notify | 001: Notify |
| |
| time | : | When you set in ‘001’ and fan motor accumulating time is | |||
|
|
| reached to 5.6 years, ‘FAn’ and PV are alternately displayed. | |||
|
|
| <Reset of accumulating time> |
| ||
|
|
| Step into | F06. Input ‘410’ and press | SET key to clear | |
|
|
|