3) Charge of refrigerant

e.

Stop vacuum pump operation.

 

Open the valve of tank to charge refrigerant

 

with approx. 0.2~0.5 Mpa/G.

 

4) Pipe pinching in High pressure side

f.

Charge refrigerant until pressure is

 

reached to 1.6MPa/G.

 

Close the valve of tank.

 

Hold charging pipe by pliers to pinch it off

 

by pinch pliers. (See picture f.)

 

Pinch pliers

5)Charge refrigerant in Low pressure side Open the valve again to charge refrigerant in Low pressure side.

Charge until you can ensure pressure should be 0.1 Mpa/G with unit running.

Turn the tank upside down to charge refrigerant without remains. (See picture-g.)

6)Pinching pipe

Measure the weight of tank after charged. Form the pipe which was pinched off.

Apply welding to the points of

pipe which were pinched to prevent being broken.

g.

h.