ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in Table 7, and also self diagnostic functions.
Table 7 Alarms and safety functions
Alarm & Safety | Situation | Indication | Buzzer | Safety operation | |
High temperature | If the chamber temperature is higher | Alarm lamp is flashed. | Intermittent tone | Remote alarm with | |
than the temperature at which the | Temperature indicator is | with 15 minutes | |||
alarm | 15 minutes delay. | ||||
high temperature alarm is activated. | flashed. | delay. | |||
|
| ||||
Low temperature | If the chamber temperature is lower | Alarm lamp is flashed. | Intermittent tone | Remote alarm with | |
than the temperature at which the | Temperature indicator is | with 15 minutes | |||
alarm | 15 minutes delay. | ||||
low temperature alarm is activated. | flashed. | delay. | |||
|
| ||||
Power failure | In the case of power failure. |
|
|
| |
When the power cord is | Alarm lamp is flashed. | Intermittent tone | Remote alarm. | ||
alarm | |||||
disconnected. |
|
|
| ||
|
|
|
| ||
| When approx. 2.8 years has passed |
|
|
| |
Battery check | with the power cord connected to | Battery check lamp lights. | |||
| the outlet. |
|
|
| |
| If the thermal sensor is | Alarm lamp is flashed. |
|
| |
| E01 and chamber temp. |
|
| ||
| disconnected.. |
|
| ||
| are displayed alternately. |
|
| ||
|
| Intermittent tone | Remote alarm. | ||
| If the thermal sensor is | Alarm lamp is flashed. | |||
|
|
| |||
| E02 and chamber temp. |
|
| ||
|
|
| |||
Sensor | are displayed alternately. |
|
| ||
|
|
| |||
abnormality | If the thermal sensor of the | Alarm lamp is flashed. |
|
| |
| E05 and chamber temp. |
|
| ||
| compressor is disconnected. |
|
| ||
| are displayed alternately. |
|
| ||
|
| Intermittent tone | Remote alarm. | ||
| If the thermal sensor of the | Alarm lamp is flashed. | |||
|
|
| |||
| E06 and chamber temp. |
|
| ||
| compressor is |
|
| ||
| are displayed alternately. |
|
| ||
|
|
|
| ||
Battery switch | When battery switch is OFF at | Alarm lamp is flashed. | |||
check | the time of alarm test. | E09 is flashed. | |||
|
| ||||
|
| Alarm lamp is flashed. |
| Remote alarm. | |
Condenser temp. | In the event of failure of fan motor |
| Compressor of | ||
E10 and chamber temp. | Intermittent tone | ||||
abnormality | for cooling the compressor. | high stage side | |||
are displayed alternately. |
| ||||
|
|
| stops. | ||
|
|
|
|
Note:
ŶAfter power failure, the operation is resumed with the condition before power failure.
ŶThe chamber temperature is displayed for 5 seconds if the alarm buzzer stop key (BUZZER) is depressed during the power failure alarm. After that, the alarm buzzer stops. The alarm lamp (ALARM) keeps blinking.
ŶIn the case of sensor abnormality (E01 or E02 is displayed), the freezer continues to run.